Shutdown & storage

The pump can be switched off under vacuum.

Short-term:

- Has the pump been exposed to conden-

sate?

Allow the pump to continue to run at atmo-

spheric pressure for a few minutes.

- Has the pump been exposed to media which

may damage the pump materials or form

deposits?

Check and clean pump heads if necessary.

- Has the pressure transducer been exposed

to media which may form deposits?

Clean pressure transducer if necessary.

Long-term:

- Take measures as described above regard-

ing short-term shutdown.

- Separate the pump from the application.

- Close inlet and outlet ports (e.g., with trans-

port caps).

- Store the pump under dry conditions.

NOTICE

NOTICE

Condensate?

O

➨ I

(5-10min)

atm