Measurement uncertainty (absolute)

after careful adjustment and at constant

temperature

<

± 0.75 Torr (1 mbar)

Temperature coefficient

<

± 0.05 Torr/K (0.07 mbar/K)

Maximum permissible temperature of

gaseous media**

continuous operation: 104°F (40°C),

for short periods (less than 5 minutes)

up to 176°F (80°C)

Ambient temperature range (operation)

50°F to 104°F (10°C to +40°C)

Ambient temperature range (storage)

14°F to 158°F (-10°C to +70°C)

Permissible relative atmospheric mois-

ture during operation (no condensation)

30% to 85%

Maximum permitted current of connect-

ed valves (connected components)

4A

Degree of protection IEC 529

(front side)

IP 42

Degree of protection IEC 529

(gauge head VSK 3000)

IP 54

Measurement connection of pressure

transducer

small flange DN 16 or

hose nozzle for tubing

I.D. 1/4” / 3/8” (6 / 10 mm)

Cable length of pressure transducer

approx. 5’6” (2m)

Supply voltage VSK 3000

(via VACUU • BUS)

6-30 V= / 5 mA

Communication VSK 3000

VACUU • BUS

Interface

RS-232 C

Dimensions CVC 3000 L x W x H

(with foot)

5.4” x 4.9” x 4.5”

(138 mm x 124 mm x 115 mm)

Dimensions VSK 3000

housing diameter

length with measurement connection

2.4”

(60 mm)

2.4”

(60 mm) with small flange

Weight CVC 3000

VSK 3000 (with small flange)

0.97 lb (440 g)

0.41 lb (185 g)

** if pumping potentially explosive atmospheres: 50 °F to 104 °F (+10°C to +40°C)