Notice – VACUUBRAND MV 10 NT VARIO User Manual

Page 40

Advertising
background image

page 40 of 126

In case of overload, the motor is shut down by

a

self-hold thermal cutout in the winding.

Note: Only manual reset is possible. Press the

START/STOP key at the controller to confirm

the error message. Switch off the pump and

disconnect from the power source. Identify and

eliminate the cause of failure. Allow the pump

to cool down sufficiently before restart.

°C / °F

I ➨ O

!

3.

2.

1.

IN

Avoid overheating (e.g., due to hot process

gases). Maximum permissible temperature

range: see ”Gas inlet temperatures”, pg. 27.

Make sure the fans’ air supply is adequate.

Do not place the pump on soft surfaces (e.g.,

rubber foam) during operation. This may re-

duce the air circulation in the pump housing

and affect the fans’ cooling effect. Check fans

regularly for dust/dirt. Clean fan guard grills if

necessary to avoid a reduction of ventilation.

air

air

A warm up period (approximately 15 min.) is

required to ensure that the rated ultimate vac-

uum and pumping speed are attained.

O

➨ I

m

3

/h

Torr/mbar

~15min

Prevent internal condensation, transfer of liq-

uids or dust. The diaphragms and valves will

be damaged, if liquids are pumped in signifi-

cant amounts.

Check the pump regularly for external soiling

and deposits. Clean the pump if necessary

to avoid an increase of the pump’s operating

temperature.

NOTICE

IN

condensate / dust

NOTICE

NOTICE

NOTICE