Connecting the pressure transducer vsk 3000 – VACUUBRAND MV 10 NT VARIO User Manual

Page 35

Advertising
background image

page 35 of 126

+

Connect a gas-tight exhaust line at the pump

outlet (see above) if necessary. Always vent

exhaust gases appropriately (e.g., into a

fume hood).

+

Never block the gas outlet. The exhaust

line must always be free of obstructions (no

back pressure) to ensure an unimpeded

discharge of gas. The cross-section of the

outlet tubing must be at least the size of the

pump’s exhaust connection.

• Reduce the transmission of vibration. Prevent mechan-

ical load due to rigid pipelines. Insert elastic hoses or

flexible elements as couplings between the pump and

rigid pipes.

The VSK 3000 is equipped with a capacitive

pressure transducer with ceramic diaphragm.

It measures the actual pressure independently

of the gas type, and with reference to the vac-

uum, i.e., absolute.

Secure hose connections at the pump ap-

propriately, e.g., with hose clamps, to protect

against accidental detachment.

EX

NOTICE

NOTICE

Connecting the pressure transducer VSK 3000

Always install outlet tubing descending from

the pump or provide other measures to avoid

backflow of condensate towards the pump.

Advertising
This manual is related to the following products: