Gas inlet temperatures – VACUUBRAND MV 10 NT VARIO User Manual

Page 27

Advertising
background image

page 27 of 126

Gas inlet temperatures

* if pumping potentially explosive atmospheres: 50 °F to 104 °F (+10°C to +40°C)

Operating condition

Inlet pressure

Permitted range of gas

temperatures at inlet

Continuous operation

> 75 Torr (100 mbar)

(high gas load)

50 °F to 104 °F

(+10°C to +40°C)

Continuous operation

< 75 Torr (100 mbar)

(low gas load)

32 °F to 140 °F*

(0°C to +60°C*)

Short-time

(< 5 minutes)

< 75 Torr (100 mbar)

(low gas load)

14 °F to 176 °F*

(-10°C to +80°C*)

** Ultimate vacuum at setting ”Pump down” at speed ”HI”

Type

ME 16 NT

VARIO

MD 12 NT

VARIO

MV 10 NT

VARIO

Maximum pumping speed

(ISO 21360)

cfm

(m

3

/h)

approx. 11.2

(approx. 19)

7.9

(13.4)

7.1

(12.1)

Ultimate vacuum

(absolute) **

Torr

(mbar)

75

(100)

1.1

(1.5)

0.2

(0.3)

Controller

CVC 3000

Pressure transducer

external gauge head VSK 3000,

ceramic diaphragm (alumina),

capacitive, absolute pressure,

gas type independent

Display

LCD graphic display, illuminated

Pressure units / scale (selectable)

Torr, mbar or hPa

Measuring range (absolute)

795 - 0.1 Torr (1060 - 0.1 mbar)

Maximum control range (absolute)*

795 - 0.1 Torr (1060 - 0.1 mbar)

Resolution (display)

0.1 Torr (mbar)

Maximum permissible pressure at

pressure transducer (absolute)

1125 Torr (1.5 bar)

* The actual vacuum control range in your application might be reduced due to ultimate vacuum

of the pump, volume of gas present, etc.