Cleaning the pressure transducer, Readjustment of the pressure transducer vsk 3000, Please continue with part 2 of this manual – VACUUBRAND MV 10 NT VARIO User Manual

Page 70

Advertising
background image

page 70 of 126

Attention: Never use a pointed or sharp-edged tool

to clean the pressure transducer.

Never touch the ceramic diaphragm of the pres-

sure transducer with hard objects.

Fill the measurement chamber with a solvent (e.g.,

benzene) and allow sufficient cleaning time. Observe

all regulations concerning usage and disposal of sol-

vents!

Drain the solvent and dispose of in accordance with

regulations. Repeat cleaning if necessary.

Rinse the measurement chamber several times with

alcohol in order to remove all solvent residues.

Allow the pressure transducer to dry.

Readjust the pressure transducer if necessary.

Readjustment of the pressure transducer VSK 3000

See section “Readjustment of VSK 3000”, pg. 67.

Cleaning the pressure transducer

Please continue with part 2 of this manual.