Connect the VSK 3000 pressure transducer

via the VACUU•BUS cable to the CVC 3000

controller.

Attention: Do not apply off-axis forces when

assembling or removing plug connections! Ob-

serve correct orientation of the plug. To con-

nect further components use VACUU•BUS Y-

adapters and extension cables.

Connect the pressure transducer via hose or

small flange connection to the vacuum appli-

cation. Do not mount the pressure transducer

directly at the pump but close to the applica-

tion. The cross-section of the tubing should be

as large as possible.

Inside a vacuum system where evaporation occurs, e.g.,

a rotary evaporator, the vacuum is not uniform. For ex-

ample, a condenser can act as a pump, or the vacuum in

the connecting tubing can be higher or lower than in the

application itself. This affects the measurement results as

well as the control levels. Therefore, carefully choose the

position where to connect the pressure transducer.

vacuum

application

(1)

(3)

(2)

Installing the hose nozzle at the pressure transducer VSK 3000: