Switch the CVC 3000 controller on.

O

➨ I

VACUU•BUS

The CVC 3000 controller uses VACUU•BUS protocol to read out the

pressure transducer via VACUU•BUS line. Maximum cable length inside

buildings: 32yd (30m). VACUU•BUS extension cable 2.2 yd (2 meters):

order-no. 612552.

The VACUUBRAND controller CVC 3000 can only be operated with com-

ponents compatible with the VACUUBRAND VACUU•BUS system, (see

“Accessories / Spare parts”, pg. 84).

The vacuum controller CVC 3000 controls VACUUBRAND diaphragm

pumps NT VARIO and optional coolant and venting valves. Connected

components (e.g., venting valve, external pressure transducer 3000 se-

ries) are automatically identified and configurated. Identical components

must be configurated beforehand; information upon request.

Do not use more than one controller within the same VACUU•BUS

system.

Several controllers in the same VACUU•BUS system will interfere

with each other and result in error messages of the connected com-

ponents (pumps, valves).

During operation

Vent and dispose of

potentially dangerous gases or

vapors at the outlet of the pump appropriately.

+

Due to the high compression ratio, the pump might

generate overpressure at the outlet. Check pressure

compatibility with system components (e.g., exhaust

tubing or exhaust valve) at the outlet. Ensure that the

pump outlet is neither blocked nor restricted.