Maximum ambient temperature: 104 °F

(40 °C)

Check the maximum temperatures, if install-

ing the pump in a cabinet or a housing.

Make sure ventilation is adequate, especial-

ly if the ambient temperature is elevated.

+

Operation with silencer at the outlet: Oper-

ating the pump at a high inlet pressure or

pumping dusty gases for a long time may

cause clogging of the silencer. Check the

silencer regularly and replace if necessary,

or install a hose nozzle instead.

• If the pump is installed at an altitude of more

than 6500 ft (2000 m) above mean sea lev-

el, check compatibility with applicable safety

requirements, and adopt suitable measures.

There is a risk of the motor overheating due

to insufficient cooling.

max.

104°F

(40°C)

• Check compatibility with the

maximally

permitted pressure at outlet and the max-

imum pressure difference between inlet

and outlet ports.

max. 6500 ft

(2000 m)

Do not start the pump if the

pressure difference between

inlet and outlet ports exceeds max. 16.0 psi (1.1 bar).

Attempts to start the pump at higher pressure difference

may cause stalling and damage of the motor.

NOTICE

max.

16.0 psi

(1.1 bar)