Technical data, General technical data valid for all pumps – VACUUBRAND MV 10 NT VARIO User Manual

Page 26

Advertising
background image

page 26 of 126

Technical data

General technical data valid for all pumps

Maximum permissible inlet pressure

(absolute)

psi

(bar)

16

(1.1)

Maximum permissible outlet pressure

(absolute)

psi

(bar)

16

(1.1)

Maximum pressure difference between

inlet and outlet

psi

(bar)

16

(1.1)

Permissible ambient temperature

storage / operation

°F

(°C)

14 to 140 / 50 to 104

(-10 to +60 / +10 to +40)

Permissible relative atmospheric mois-

ture during operation (no condensation)

%

30 to 85

Maximum permissible installation alti-

tude above mean sea level

ft

(m)

6500

(2000)

Rated motor power

hp (kW)

0.71 (0.530)

No-load speed

rpm

30 - 2400

Maximum permissible range

of supply voltage (±10%)

Attention: Observe specifications of rating plate!

100-120 V~ 50/60 Hz,

200-230 V~ 50/60 Hz

Maximum rated current at:

100-120 V~ 50/60 Hz

200-230 V~ 50/60 Hz

A

A

8.0

3.5

Device fuse

2 slow blow fuses 8A

Motor protection

thermal cutout, manual reset

Degree of protection IEC 529

IP 40

A-weighted emission sound pressure

level* (uncertainty K

pA

: 3 dB(A))

dB(A)

50

Inlet

small flange KF DN 25

Outlet

silencer / thread G 1/2”

Dimensions L x W x H

approx.

in

(mm)

21.8 x 10.2 x 16.5

(554 x 260 x 420)

Weight approx.

lbs. (kg)

68.8 (31.2)

* Measurement according to EN ISO 2151:2004 and EN ISO 3744:1995 at 1500rpm and ulti-

mate vacuum with standard silencer at outlet.

Advertising
This manual is related to the following products: