VACUUBRAND MV 10 NT VARIO User Manual

Page 18

Advertising
background image

page 18 of 126

Pumps without the

`

mark on the rating plate are

not approved for the pumping of potentially explo-

sive atmospheres. Do not pump potentially explo-

sive atmospheres with those pumps

Pumps bearing the

`

mark on their rating plates

are

approved for the pumping of potentially explo-

sive atmospheres according to their classification

II 3G IIC T3 X according to ATEX, but they are not

approved for operation in potentially explosive at-

mospheres (see section “

`

Important information:

Equipment marking (ATEX)”, pg. 24)

.

The pumps are

not suitable to pump any of the sub-

stances listed below.

Do not pump:

-

unstable substances

- substances which react explosively under

impact

(mechanical stress) without air

- substances which react explosively when being ex-

posed to

elevated temperatures without air,

- substances subject to

auto-ignition,

- substances which are

inflammable without air

-

explosive substances.

The pumps are

not approved for operation below

ground.

Do not operate the pump below ground.

Do not pump substances which may form depos-

its inside the pump. The pumps are not suitable for

pumping substances which may form deposits inside

the pump. Deposits and condensate in the pump may

lead to increased temperatures even to the point of

exceeding the maximum permitted temperatures.

• Check the inlet and outlet of the pump, if there is a

danger of forming

deposits inside the pump, e.g., in

the pump chambers (the pump chamber is the part

between diaphragm and head cover. See section “Re-

placing diaphragms and valves”, pg. 91). Inspect the

pump chambers regularly and clean if necessary.

Advertising
This manual is related to the following products: