• Pay attention to the safety symbol ”hot surfaces” on

the equipment. Hot parts may cause burns if touched.

Adopt suitable measures to prevent any danger aris-

ing from hot surfaces or electric sparks. Ensure that

hot surfaces of the pump do not cause burns. Provide

a suitable contact guard if necessary.

• Ensure that the

coolant outlet pipe at the waste vapor

condenser is always free and that it cannot get blocked.

Do not start the pump if the pressure difference between

inlet and outlet exceeds 16 psi (1.1 bar) at maximum.

Prevent the backpressure of gases and the backflow of

condensates at the outlet.

Check the liquid level in the catchpots regularly and drain

condensate in time to prevent overfilling.

Provide appropriate protective measures to allow for the

possibility of failure and

malfunction. The protective mea-

sures must also allow for the requirements of the respec-

tive application.

Pumps with AC motor:

In case of overload, the motor is shut down by a

self-hold

thermal cutout in the winding.

Note: Only manual reset is possible. Switch off the pump

and disconnect from the power source. Identify and elimi-

nate the cause of failure. Wait approx. five minutes before

restarting the pump.

Note: In case of supply voltage below 100V, the lock

of the cutout might be impaired and the pump may

restart on its own after sufficient cooling down. Take

appropriate precautions, if an automatic restart of the

pump may lead to a dangerous situation (e.g., switch

off the pump and disconnect from the power source).