Connect a gas-tight exhaust line at the pump

outlet if necessary. Always vent exhaust

gases appropriately (e.g., into a fume hood).

+

Never block the gas outlet. The exhaust

line must always be free of obstructions (no

back pressure) to ensure an unimpeded

discharge of gas. The cross-section of the

outlet tubing must be at least the size of the

pump’s exhaust connection.

• Reduce the transmission of vibration. Prevent mechan-

ical load due to rigid pipelines. Insert elastic hoses or

flexible elements as couplings between the pump and

rigid pipes.

Secure hose connections at the pump ap-

propriately, e.g., with hose clamps, to protect

against accidental detachment.

To reduce pump noise emanating from the

pump exhaust port, connect an exhaust hose

or use a silencer (see „Accessories“, pg. 54).

NOTICE

EX

NOTICE

Exhaust waste vapor condenser (EK) at the outlet

The

exhaust waste vapor condenser enables an efficient condensation

of the pumped vapors at the outlet.
+

No backflow of condensates.

+

Controlled recovery of condensates.

+

Close to 100% solvent recovery.

+

The isolation cover protects against glass splinters in case of break-

age, acts as thermal isolation to avoid condensation of humidity and is

intended to absorb shocks.

Always install outlet tubing descending from

the pump or provide other measures to avoid

backflow of condensate towards the pump.