Electrical connection, Notice – VACUUBRAND MD 1C VARIO-SP User Manual

Page 41

Advertising
background image

page 41 of 87

• Only install the optional coolant valve in the

supply line of the exhaust waste vapor con-

denser.

Note: Install the hoses of the cooling system

in a way to avoid the flow / dripping of con-

densed water onto the pumping unit (espe-

cially cables and electronic parts, see also

IP degree of protection, „Technical data“,

pg. 25).

• Ensure that the

coolant outlet tubing is always unob-

structed and that it cannot get blocked.

• Maximum permissible coolant pressure at the exhaust

waste vapor condenser: 87 psi (6 bar) absolute. Outlet

flow must always be unhindered.

• Comply with the maximum permissible coolant pres-

sures of additional components in the coolant circuit

(e.g., coolant valve).

• Avoid overpressure in the coolant circuit (e.g., caused

by blocked or squeezed coolant hoses).

EX

IN

Permissible range of coolant temperature at the exhaust

waste vapor condenser: 5°F to 68°F (-15°C to +20°C)

Check hose connections prior to starting operation of the

cooling system.

Check coolant hoses regularly during operation.

NOTICE

MD 1C (AC motor)

Plug in the power cord.

• Check the power source and the pump’s

rating plate to be sure that the power source

and the equipment match in voltage, phase,

and frequency.

Switch the pump on.

Electrical connection