Voltage selection switch:

1. Disconnect the electrical

power cord.

2. Use a screw driver to ad-

just the

voltage selection

switch to the supply volt-

age:

• Check that the voltage selection switch is posi-

tioned correctly. Check every time before starting

the pump.

Note: If the pump is switched on with wrong volt-

age selection, the motor may be damaged!

Change the selection at the voltage selection

switch only if the pump is unplugged from the

power source.

Only pumps with dual-voltage motor:

”115/120” corresponds to 100-120 V and

”230/240” corresponds to 200-230 V.