In case of condensation

In case of

condensation: Check the liquid lev-

el in the catchpots during operation. Check the

liquid level in the catchpots regularly. Do not

allow the catchpots to overfill. Drain catchpots

in time to avoid overflow. Install a level sensor

(see „Accessories“, pg. 54) for monitoring,

if necessary (VACUUBRAND controller CVC

3000 or VNC 2 is required).

The maximum liquid level is at approximately

80% of the total filling level to avoid problems

when removing the catchpots.

Drain catchpots.

+

Important: Comply with regulations when

disposing of solvents/condensates. Recy-

cle if possible; purify if contaminated.

Removing the catchpots:

Catchpot at outlet:

Remove joint clip. Remove catchpot and drain

condensate.

Catchpot at inlet:

Admit air or inert gas (via the pump inlet) to

restore atmospheric pressure in the catchpot

before attempting removal. Remove joint clip.

Remove catchpot and drain condensate.

Reattach drained catchpots.