Pumps with DC motor:

A temperature sensor at the circuit board protects the mo-

tor: Current limitation in case the temperature at the circuit

board raises above 158°F (70°C). At temperatures above

185°F (85°C) the pump switches off. In case of a motor

blockage (after 10 start-up attempts) the pump switches

off.

Note: Only manual reset is possible. Disconnect the pump

from the power source. Identify and eliminate the cause of

failure.

NOTICE

Note: In case of supply voltage below 100V, the lock

of the breaker may not latch and the pump might re-

start on its own after sufficient cooling. Take appropri-

ate precautions, if an automatic restart of the pump

may lead to a dangerous situation.

Make sure the fan’s air supply is adequate.

Check fan regularly for dust/dirt. Clean fan

guard grill if necessary to avoid a reduction of

ventilation.

NOTICE

A warm up period (approximately 15 min.) is

required to ensure that the rated ultimate vac-

uum, the pumping speed, and the full vapor

pumping rate are attained.

O

➨ I

m

3

/h

Torr/mbar

~15min

air

Important notes regarding the use of gas ballast

Gas ballast is a continuous purge to keep the pump’s in-

terior as clean as possible and to reduce the possibility of

condensation inside the pump.