During operation – VACUUBRAND MD 1C VARIO-SP User Manual

Page 48

Advertising
background image

page 48 of 87

During operation

Vent and dispose of

potentially dangerous gases or

vapors at the outlet of the pump appropriately.

+

Due to the high compression ratio, the pump might

generate overpressure at the outlet. Check pressure

compatibility with system components (e.g., exhaust

tubing or exhaust valve) at the outlet. Ensure that the

pump outlet is neither blocked nor restricted.

+

Operation with silencer (see „Accessories“, pg. 54)

at the outlet: Operating the pump at a high inlet pres-

sure or pumping dusty gases for a long time may cause

clogging of the silencer. Check the silencer regularly

and replace if necessary, or install a hose nozzle in-

stead.

+

Maximum ambient temperature: 104 °F

(40 °C)

Check the maximum temperatures, if install-

ing the pump in a cabinet or a housing.

Make sure ventilation is adequate, especial-

ly if the ambient temperature is elevated.

max.

104°F

(40°C)

• If the pump is installed at an altitude of more

than 6500 ft (2000 m) above mean sea lev-

el, check compatibility with applicable safety

requirements, and adopt suitable measures.

There is a risk of the motor overheating due

to insufficient cooling.

max. 6500 ft

(2000 m)

• Check compatibility with the

maximally

permitted pressure at inlet and at outlet

and the

maximum pressure difference

between inlet and outlet ports.

max.

16.0 psi

(1.1 bar)

absolute