Do not start the pump if the

pressure difference between

inlet and outlet ports exceeds max. 16.0 psi (1.1 bar).

Attempts to start the pump at higher pressure difference

may cause stalling and damage of the motor.

NOTICE

gas ballast

NOTICE

If pumping condensable vapors (water vapor,

solvents, etc.), let the pump run with

gas bal-

last to help purge any condensation in the

pump.

Prevent internal condensation, transfer of liq-

uids or dust. The diaphragms and valves will

be damaged, if liquids are pumped in signifi-

cant amounts.

Check the pump regularly for external soiling

and deposits. Clean the pump if necessary

to avoid an increase of the pump’s operating

temperature.

NOTICE

IN

condensate / dust

Avoid overheating (e.g., due to hot process

gases). Maximum permissible temperature

range: see „Gas inlet temperatures“, pg. 27.

NOTICE

Pumps with AC motor:

In case of overload, the motor is shut down by

a

self-hold thermal cutout in the winding.

Note: Only manual reset is possible. Switch

off the pump and disconnect from the power

source. Identify and eliminate the cause of fail-

ure. Allow the pump to cool down sufficiently

before restart.

°C / °F

I ➨ O

!

3.

2.

1.

NOTICE