Replacing diaphragms and valves

+

Please read section ”Replacing diaphragms and valves” com-

pletely before starting maintenance.

The pictures may show other versions of pumps. This does not change

the method of replacing diaphragms and valves.

Never operate the pump if covers or other parts of

the pump are disassembled.

Before starting maintenance,

disconnect the elec-

trical power cord (MD 1C with AC motor), respec-

tively the control line (MD 1C VARIO-SP). Wait five

seconds after isolating the equipment from AC power

or from supply voltage to allow the capacitors to dis-

charge.

Ensure that the pump cannot be operated accidentally.

Note: The pump might be contaminated with the pro-

cess chemicals that have been pumped during opera-

tion. Ensure that the pump is decontaminated before

maintenance.

Avoid the release of pollutants.

+

Never operate a defective or damaged pump.

+

Check every motor capacitor (MD 1C with AC motor)

regularly by measuring its capacity and estimating its

service life. Replace old capacitors early enough to

prevent a failure. The capacitors must be replaced by

a trained electrician.

+

Take adequate precautions to protect people

from the effects of dangerous substances

that may have contaminated the pump and

may be released upon disassembly. Ensure

that the maintenance technician is familiar

with the safety procedures which relate to the

products processed by the pumping system.

Use appropriate protective clothing, safety

goggles and protective gloves.