Catchpot:

The catchpot is coated with a protective layer

to protect against shattering in case of break-

age or implosion.

Assemble the catchpot at the exhaust waste

vapor condenser using a joint clip.

over-

pressure

safety

relief

device

+

Check the overpressure safety relief device

at the exhaust waste vapor condenser regu-

larly; replace if necessary. Check especially

for deterioration, coalescence and cracks.

Assemble the hose nozzles for coolant in-

let and coolant outlet tubing at the exhaust

waste vapor condenser (hose nozzles for

tubing I.D. 1/4”-5/16” (6-8 mm)).

Attach the tubing of the coolant circuit to the

respective hose nozzles at the waste vapor

condenser.

Prevent the discharge of dangerous gases and vapors

to the surrounding atmosphere. If appropriate, connect

the exhaust line to a suitable treatment system.

Check all hose connections prior to starting

operation of the cooling system.

Secure coolant hoses at the hose nozzles

(e.g., with hose clamps) to prevent their acci-

dentally slipping off.

EX

IN

+

Never block the gas outlet (hose nozzle for

tubing I.D. 3/8” (10 mm)). The exhaust hose

has always to be unobstructed and with-

out back pressure to enable an unhindered

discharge of gases and protect the pump

valves from damage.

NOTICE

outlet