Notice – VACUUBRAND MD 1C VARIO-SP User Manual

Page 37

Advertising
background image

page 37 of 87

• Reduce the transmission of vibration. Pre-

vent mechanical load due to rigid pipelines.

Insert elastic hoses or flexible elements

as couplings between the pump and rigid

pipes.

Note: Flexible elements will compress or

flatten when evacuated if not designed for

use under vacuum.

• Hose connections at the pump inlet must al-

ways be gas tight.

MD 1C + AK + EK:

Assembling the hose nozzle with union nut:

Take the hose nozzle with attached com-

pression ferrule and union nut out of the

catchpot and put onto inlet connection.

Tighten the union nut by hand until you can

feel the stop. Then tighten an additional 1/4

rotation with an open-ended wrench (size

17mm) for final installation.

Use connecting hoses with large diameter and

keep them as short as possible to avoid flow

losses. Locate the pump as closely as possi-

ble to the application.

NOTICE

Secure hose connections at the pump ap-

propriately, e.g., with hose clamps, to protect

against accidental detachment.

NOTICE

+

Particles and dust must not be aspirated. If necessary,

you must install appropriate filters. You must ensure

their suitability concerning gas flow, chemical resis-

tance and resistance to clogging prior to use.

Advertising
This manual is related to the following products: