Heath Consultants GT Series User Manual

Page 106

Advertising
background image

11-2

USER HANDBOOK

Note: A Time Weighted Average (TWA) value is the mean

average gas level over a specific period. The STEL

is 15 minutes and the LTEL is 8 hours. In

accordance with legislation, this requires the time

weighted averages to be averaged over a full

period whether the instrument is ON or OFF. Such

averaging essentially makes the instrument single

user applicable. The option is available to restart

the averaging after each instrument switch-off,

thus allowing for multiple user application.
A TWA alarm is intermittent with the actual gas

reading, therefore it is possible to get a value of

zero (0) and an LTEL or STEL alarm.

All alarms are user configurable to meet the specific needs

of different companies.

Note: The gas alarm levels, instantaneous STEL and LTEL,

are set at the time of instrument manufacture. It is

important that the user ensures that the levels are

in accordance with their company’s alarm levels

and with health and safety legislation. The alarm

levels may be changed, if required, as detailed in

the ‘SET-UP SOFTWARE USER HANDBOOK’.