Safety – Heath Consultants GT Series User Manual

Page 4

Advertising
background image

ii

USER HANDBOOK

SAFETY
•

The instrument must be regularly serviced and calibrated by fully trained

personnel in a safe area.

•

Batteries: Alkaline or *Rechargeable batteries must be exchanged (*and

recharged) in a safe area and fitted correctly before use.

Never use damaged batteries or expose to extreme heat.

See Chapter 12 : OPERATOR MAINTENANCE.

•

Only GMI replacement parts should be used.

•

If the instrument detects gas, follow your own organisation’s procedures

and operational guidelines.

•

The combustion chamber is a flameproof assembly and must not be opened

in the presence of a flammable atmosphere.

•

GT series instruments are certified as ATEX

II 2 G EEx iad IIB T3

(-20oC < Tamb < 50oC).

UL Class 1 Groups C and D.

•

This equipment is designed and manufactured to protect against other

hazards as defined in paragraph 1.2.7 of Annex II of the ATEX Directive

94/9/EC

Any right of claim relating to product liability or consequential damage to any third

party against GMI is removed if the warnings are not observed.
WARNING: To prevent ignition of flammable or combustible atmospheres, remove

batteries before servicing.
WARNING: To prevent ignition of flammable or combustible atmospheres, read,

understand and adhere to the manufacturer’s live maintenance procedures.
WARNING: To reduce the risk of ignition of a flammable or explosive atmosphere,

batteries must be changed only in a location known to be non-hazardous.
WARNING: To reduce the risk of explosion, do not mix old batteries with used

batteries or mix batteries from different manufacturers.
WARNING: Never attempt to recharge non rechargeable cells.
CAUTION: Not for use in oxygen enriched atmospheres.
CAUTION: Replace batteries only with approved batteries:

NiMH – GP350
Alkaline – Duracell, Energizer, Rayovac.

Advertising