Table. 11.1 alarm indication – Heath Consultants GT Series User Manual

Page 115

Advertising
background image

11-11

ALARMS

ALARM TYPE

LATCHING

Yes / No

MUTE

Yes / No

AUDIBLE

INDICATION

VISUAL LED

INDICATION

DISPLAY

LEL Warning

N

N / A

N / A

N / A

Flash Range

LEL (

HI

)

N

Y

High Pitch

Tone

(4) Flashing

Toggle

Hi / Conc

LEL (

HIHI

)

Y

N

High Pitch

Warble

(8) Ramping

Toggle

Hi Hi / Conc

O

2

(

LO

)

N

Y

High Pitch

Tone

(4) Flashing

Toggle

Lo / Conc

O

2

(

LOLO

)

Y

N

High Pitch

Warble

(8) Ramping

Toggle

Lo Lo / Conc

O

2

(

HIHI

)

Y

N

High Pitch

Warble

(8) Ramping

Toggle

Hi Hi / Conc

H

2

S Warning

N

N / A

N / A

N / A

Flash Range

H

2

S (

HIHI

)

Y

N

High Pitch

Warble

(8) Ramping

Toggle

Hi Hi / Conc

H

2

S (STEL)

Y

N

High Pitch

Warble

(8) Ramping

Toggle

STEL / Conc

H

2

S (LTEL / TWA)

Y

N

High Pitch

Warble

(8) Ramping

Toggle

LTEL / Conc

CO Warning

N

N / A

N / A

N / A

Flash Range

CO (

HIHI

)

Y

N

High Pitch

Warble

(8) Ramping

Toggle

Hi Hi / Conc

CO (STEL)

Y

N

High Pitch

Warble

(8) Ramping

Toggle

STEL / Conc

CO (LTEL / TWA)

Y

N

High Pitch

Warble

(8) Ramping

Toggle

LTEL / Conc

LO

/

Y

N / A

Low Pitch

Tone

Fault LED

Flashing

Fault 'Term'

Low Battery

Y

N / A

Low Pitch

Tone

Fault LED

On

Fault 'Term'

Zero Fault

Y

N / A

Low Pitch

Tone

Fault LED

Flashing

Sensor Fault

Y

N / A

Low Pitch

Tone

Fault LED

Flashing

Flow Fault

Y

N / A

Low Pitch

Tone

Fault LED

Flashing

Calibration
Expired

Y

N / A

Low Pitch

Tone

Fault LED

Flashing

Service Expired

Y

N / A

Low Pitch

Tone

Fault LED

Flashing

Fault 'Term'

Fault 'Term'

Fault 'Term'

Fault 'Term'

Fault 'Term'

Table. 11.1 Alarm Indication