Heath Consultants GT Series User Manual

Page 128

Advertising
background image

12-4

USER HANDBOOK

4) Inspect the filter for contamination or damage.

If either is evident, replace the filter.

5) Fit a new Hydrophobic Probe Filter (Part No.

67259), if required.
Note:

Installation is a reversal of the removal

procedure and therefore relevant Figs.

should be referred to.

6) Position hydrophobic probe filter so that slots in

filter locate corectly over lugs in probe tube then

push filter until correctly seated. Care must be

taken not to damage the figaro sensor.

7) Hold probe tube then attach probe filter bulb to the

flexible probe assembly, in a clockwise direction,

and tighten to secure.
Note:

Care must be taken not to overtighten the

probe filter bulb.

8) Before attaching the probe connector to the probe

filter bulb, check the probe washer for signs of

damage. If damaged, it must be replaced with a

new probe washer (Part No.12379).

9) Hold probe filter bulb then attach probe connector

to the probe filter bulb, in a clockwise direction,

and tighten to secure.
Note:

Care must be taken not to overtighten the

probe connector.

C12.pmd

22/06/2006, 11:43

4