OPERATING PROCEDURE

OPERATING PROCEDURE

Check the following:

•

The instrument is clean and in good condition.

•

The batteries are in good condition, fully charged

and fitted correctly.

•

The hydrophobic filter is clean and in good condi-

tion.

•

The sample line and any other accessories used

are in good condition.

•

The ‘clear bulb’ at probe end is screwed tightly.

•

Switch instrument ON in fresh air.

•

The battery indication provides sufficient

capacity for the operation.

•

The instrument is within the calibration period you

have decided is necessary for your application.

•

All applicable ranges are operational.

•

There are no fault indications.

•

Attach optional accessories, as required.

•

The instrument displays a sample / flow fault when

inlet is blocked (with a finger for example). If fault

is not displayed, check tightness of all fittings.

•

After use, allow the instrument to run for 1 – 2

minutes in fresh air before switching the instru-

ment OFF.