5

Pre Installation

Installation Requirements

Feedwater
Requirements

Type of Feedwater

Potable

Flowrate

≥ 10.7 GPH (40 LPH)

Minimum Pressure

≥ 7.3psi (0.5 bar)

Maximum Pressure

≤ 87.0 psi (6 bar)

Conductivity

< 2000 µS/cm

Temperature

5 to 35 °C

pH

4 - 10

Fouling Index

< 10

Iron

< 0.1 ppm as CaCO

3

Aluminum

< 0.05 ppm as CaCO

3

Manganese

< 0.05 ppm as CaCO

3

Free Chlorine

< 1 ppm

Langelier Saturation Index

< + 0.2

TOC

< 2000 ppb

Feedwater
Connection Needed

Feedwater Piping Connection

1/2 inch Male GAZ, NPTM or BSPM

Reject Flow
Requirement

Drain Capacity

7.9 GPH (30 LPH)

Environmental

Requirements

Indoor Use Only

Storage Temperature

5 °C < T < 40 °C

Operating Temperature 5 °C < T < 40 °C

Relative Humidity

Should not exceed 80% for temperatures below 31 °C
Should stay within 50% to 80% between 31 °C and
40 °C.

Altitude

< 9843 ft (3000 metres)

Installation Category

II

Pollution Degree

2