Labconco WaterPr BT Systems 90153 Series User Manual

Page 55

Advertising
background image

Maintenance

47

4

Open the Tank Outlet Valve. Allow the tank
to be emptied to its lowest water level.

NOTE: The lowest water level is not

calibrated if the amount of water emptied

from the tank is less than 10%.

5

Close the Tank Outlet Valve.

6

Press the “+” and “-“ Buttons together to
enter the lowest water level. The Display will
exit the menu option and go into FILLING
TANK mode.

The Tank Level is now calibrated.

NOTE: If the Display remains in the menu option and the highest tank level is

blinking, then the tank level was not calibrated. The amount of water emptied

from the tank was not enough (less than 10%). The tank needs to be emptied

to its lowest water level. Allow the tank to fill to the 100% level and perform

the calibration again.