Installation – Labconco WaterPr BT Systems 90153 Series User Manual

Page 30

Advertising
background image

Installation

22

4

Measure the total volume of water (in
Liters) dispensed from the system using a
1 L graduated cylinder.

5

Press the “+” or “-“ Button to match the
volume Display to the volume measured.

6

To exit the menu, press and hold the
Main Button for 2 seconds.

To display the next menu option, press
the Main Button once.

The Product Water flowrate is now
calibrated.

Advertising
This manual is related to the following products: