Flush mode – Labconco WaterPr BT Systems 90153 Series User Manual

Page 49

Advertising
background image

Maintenance

41

Flush Mode

ATTENTION

The Tank Outlet Valve should be left open during FLUSH mode.

H

The system will now go into FLUSH mode for 15 minutes (H). This is

done to empty the Filter Pack of air and hydrate the material inside.

I

When FLUSH mode is finished, the system will go into FILLING TANK

mode automatically (I).

Rinsing the Filter Pack

J

Let FILLING TANK mode run for a minimum of 2 hours with the Tank

Outlet Valve open and placed to the drain. This will completely rinse
the purification media inside the Filter Pack.

NOTE: for sensitive applications, it is recommended to leave the system

in FILLING TANK overnight to ensure complete rinsing of the RO

membrane.

Close the Tank Outlet Valve (J) and connect it back to the system if

needed. The tank will start to fill up with water. It could take
approximately 1.5 to 2 hours to fill the tank. When it is full, the
Display will indicate a full tank by illuminating the symbols
representing the Tank Level

(see Section Understanding the Display)

.

The Vent Filter needs to be installed. See

Section Installation of the

Vent Filter

.