How to Calibrate the Flowrate from the WaterPro® BT System (F02)

Before calibrating the Product flowrate from the WaterPro® BT System, you will need a 1 L graduated cylinder to
measure the total volume of water that will be dispensed. The Final Filter should be installed.

1

Press the Main and “-“ Buttons together
to enter the menu. The Display will show
F01.

2

Press the Main Button once. The Display
will show F02 and the 60 second timer.
The graduated cylinder will be blinking.

3

Press the Dispensing Button once. The
system will dispense water for 60 seconds.
The Display will show the timer counting
down.

NOTE: If you press the Dispensing

Button again or if the Tank Level

display is at the 10% level before 60

seconds have elapsed, then the system

will stop dispensing water. Calibration

of the flowrate is cancelled.

Continued on next page