Troubleshooting

64

Problem

Possible causes

What to do

The last Product resistivity
value is not displayed when
you press the “-“ button in
FILLING TANK mode or in
PRE OPERATE mode.

A measurement was not made

during DISPENSING mode or during
recirculation.

The value is out of measurement

range.

Dispense or recirculate water

manually to start a Product
resistivity reading again.

The Tank Outlet Valve is not open.

The system detects that there is
water in the tank. The system will
not resume FLUSH mode until the

tank is emptied of water.

Open the Tank Outlet Valve to

drain the water from the tank.

Check that the Tank Outlet Valve

and Tubing are directed downwards
into the drain.

The Filter Pack lifetime is

exhausted.

The Product resistivity is less than

the Resistivity Setpoint set in menu
option C02.

Replace the Filter Pack. See

Section

How to Replace the Filter Pack

Cartridge

.

The Filter Pack is not installed

correctly or not installed at all.

Reinstall the Filter Pack.

If the red Pack Alarm is still

displayed, then contact your

technical support.

The system may have been in

STANDBY mode for a while.

The Product resistivity is below the

Resistivity Setpoint set in menu
option C02.

Recirculate the water in the

system. See

Section How to

recirculate water manually before

dispensing.

Replace the Filter Pack. See

Section

How to Replace the Filter Pack

Cartridge

.