How to calibrate the tank level (c04), Ion how to calibrate the tank level (c04) – Labconco WaterPr BT Systems 90153 Series User Manual

Page 54

Advertising
background image

Maintenance

46

How to Calibrate the Tank Level (C04)

Before calibrating the tank level, the tank needs to be filled to the 100% level or TANK FULL.

1

Press the Main and “+“ Buttons together to
enter the menu. The Display will show C01.

2

Press the Main Button 3 times. The Display
will show C04. The highest Tank Level display
will be blink ing. This means that the highest
water level is ready to be calibrated.

3

Press the “+” and “-“ Buttons together to
enter the highest water level. The Display
will show the lowest Tank Level (red level)
display blinking.

NOTE: The highest and lowest tank level

must be calibrated together or else the

tank level is not calibrated.

Continued on next page