Maintenance

61

How to View or Reset the UV Lamp Timer (C05)

The UV Lamp Timer should be reset only after the UV Lamp has been replaced (see

Section How to Replace the UV

Lamp (UV System only))

. The UV Lamp Timer displays the time left until the UV Lamp needs to be replaced. The

Display will show the red UV Lamp Alarm icon blinking when the Timer reaches 0 days. This message is displayed
until the UV Lamp is replaced and the UV Lamp Timer is reset.

How to View the Remaining Days on the UV Lamp Timer

1

Press the Main and “+“ Buttons together to
enter the menu. The Display will show C01.

2

Press the Main Button 4 times. The Display will
show C05 and the days left on the UV Timer.

3

To exit menu, press and hold the Main Button
for 2 seconds.

To display the next menu option, press the
Main Button once.