background image

RVC 2-33 CD

&

Set-up and Connection

24

Version 01/2009, Rev. 1.21 of 25/11/2013 - sb

Translation of the original operating manual

5.4 Power Supply

5.4.1 Connection


The operating voltage on the name plate must correspond to the local
supply voltage!


christ

rotational vacuum concentrators are units of safety

class I and have a three-wire power cord with a IEC C13 connector
(see also chapter 10 "Technical data").
At the rear panel of the rotational vacuum concentrator, there is an
equipotential bonding screw (see fig. 2.2, page 11) for the ground
wire check.

5.4.2 Fuses

Typically, the rotational vacuum concentrators must be protected on-
site with 16 Amp G fuses.

5.4.3 Connection of Cooling Trap and/or Vacuum Pump

The vapours that are formed can be withdrawn and condensed in the
following ways by the rotational vacuum concentrator

RVC 2-33 CD

&

:


Withdrawal of the vapours by a vacuum pump, e.g. the vacuum
diaphragm pump for chemical applications "MZ 2C" or "MD 4C",
followed by condensation in a liquid-cooled emission condenser.
For low-boiling samples containing solvents:

1 Vacuum pump

2 Vacuum hose

3 Stop valve

4 Rotational vacuum

concentrator

Fig. 5.1: Connection of the vacuum pump
Attention: The set-up in the photo is for demonstration purposes only!

Normally, the stop valve is installed behind the rotational vacuum

concentrator and cannot be seen from the front.


The Vacuum pump must be connected to the rotational vacuum
concentrator as shown in the figure above. The connector of the stop
valve must be plugged into the socket on the back of the unit.

1 2

3

4