Martin Christ RVC 2-33 CDplus User Manual

Page 36

Advertising
background image

RVC 2-33 CD

&

Using the RVC

36

Version 01/2009, Rev. 1.21 of 25/11/2013 - sb

Translation of the original operating manual

The available measurement values can be displayed in the values
window on the left as well as in the values window on the right so that
a combination of the following values can be selected:

Rotor speed
Vacuum (only if a vacuum sensor is installed)

Safety pressure (only if a vacuum sensor is installed)

Rotor chamber temperature

Ice condenser temperature (only if the option “Cooling trap
control” is used)

Total time (counts the time of the entire process run)

Section time (counts the time of an individual phase, e.g.warm-
up, program sections)


6.3.4 Mode

The mode selection can be activated by pressing the left-hand
function key "mode" in the active values window. The following
phases are available:

Open lid
Warm Up
Evaporation manual
Evaporation program


6.3.4.1

Opening the Lid

The lid can only be opened if the rotor is at a standstill. It is not
possible to open the lid during an evaporation process.

Press the left-hand function key "mode"

Select "Open lid" with the up and down keys

Press the right-hand key "enter" to confirm

Fig. 6.8: Selecting "Open lid"

During the evaporation process, the housing, the lid and the interior of
the rotational vacuum concentrator can reach surface temperatures of
more than +50°C.
Risk of burns!