3 evaporation summary – Martin Christ RVC 2-33 CDplus User Manual

Page 73

Advertising
background image

RVC 2-33 CD

&

Appendix

Version 01/2009, Rev. 1.21 of 25/11/2013 - sb

73

Translation of the original operating manual

11.3 Evaporation Summary

Vials

Number of

samples

Sample volume

(ml)

Temperature (°C)

Pump

Cooling trap

CT 02-50

Total time (min)

Water

100 ml test tube

12

50

30

MZ 2C

yes

1300

100 ml test tube

12

50

45

MZ 2C

yes

910

100 ml test tube

12

50

60

MZ 2C

yes

810

100 ml test tube

12

50

80

MZ 2C

yes

615

Acetonitrile C

2

H

3

N

100 ml test tube

12

50

30

MZ 2C

yes

230

100 ml test tube

12

50

45

MZ 2C

yes

160

100 ml test tube

12

50

60

MZ 2C

yes

160

100 ml test tube

36

1

30

MZ 2C

yes

90

1,4 Dioxane C

4

H

8

O

2

100 ml test tube

12

50

30

MZ 2C

yes

270

100 ml test tube

12

50

45

MZ 2C

yes

225

100 ml test tube

12

50

60

MZ 2C

yes

155

100 ml test tube

12

50

80

MZ 2C

yes

110

Toluol C

6

H

5

CH

3

100 ml test tube

12

50

30

MZ 2C

yes

165

100 ml test tube

12

50

45

MZ 2C

yes

150

100 ml test tube

12

50

60

MZ 2C

yes

110

100 ml test tube

12

50

80

MZ 2C

yes

95

Methanol CH

3

OH

100 ml test tube

12

50

30

MZ 2C

yes

285

100 ml test tube

12

50

45

MZ 2C

yes

210

100 ml test tube

12

50

60

MZ 2C

yes

195

100 ml test tube

12

50

80

MZ 2C

yes

180

Ethanol C

2

H

6

O

100 ml test tube

12

50

30

MZ 2C

yes

235

100 ml test tube

12

50

45

MZ 2C

yes

197

100 ml test tube

12

50

60

MZ 2C

yes

180

100 ml test tube

12

50

80

MZ 2C

yes

145

Advertising