Malfunctions and Error Correction

50

Version 01/2009, Rev. 1.21 of 25/11/2013 - sb

Translation of the original operating manual

7 Malfunctions and Error Correction

7.1 General Malfunctions

Malfunctions are indicated in the "Process & equipment information
system" (see 6.3.5.4). An acoustic signal sounds when an error
message is generated.

Eliminate the source of the problem (see 7.1.1 to 7.1.4).
Acknowledge the error messages by pressing the right-hand

function key.

7.1.1 Error Messages

Type of error

Possible reason

Correction

No vacuum sensor signal

The control unit receives an invalid
value from the vacuum sensor

Check the connecting cables
and plug-in connections of the
vacuum sensor. If necessary,
replace the vacuum sensor.

No rotor chamber sensor
signal

The temperature sensor of the
rotor chamber is disconnected or
defective.

Contact the after-sales service
(see 7.2).

No rotor sensor signal

The rotor temperature sensor is
disconnected or defective

Contact the after-sales service
(see 7.2).

No ice condenser sensor
signal

The data communication link to
the cooling trap is interrupted, or
the ice condenser temperature
sensor is disconnected or
defective.

Check the data link between the
rotational vacuum concentrator
and the cooling trap.

Lid cannot be opened

The rotor chamber is not aerated
completely.

Aerate the rotor chamber
completely.

Lid contacts error

the system has detected an open lid
during evaporation.The rotor will be
stopped immediately. The lid must
be closed while the rotor is moving.

Check the lid contacts. If
necessary, contact the after-sales
service (see 7.2).

Rotor overtemperature

The maximum permissible rotor
temperature of 85°C has been
exceeded.

If the system does not cool down,
disconnect it immediately from
the power supply. Contact the
after-sales service (see 7.2).

Chamber overtemperature

The maximum permissible chamber
temperature of 85°C has been
exceeded.

If the system does not cool down,
disconnect it immediately from
the power supply. Contact the
after-sales service (see 7.2).

Permissible imbalance
exceeded

The rotor will be stopped
immediately, the process will also
be stopped.
- improper loading of the rotor
- glass breakage
- uneven evaporation

Balance the load and restart the
rotational vacuum concentrator. If
the error occurs again, contact
the after-sales service (see 7.2).

VSK3000 calibration error

Incorrect calibration of the vacuum
sensor VSK3000.

The sensor must be recalibrated.