11 appendix, 1 mathematical relations, Appendix – Martin Christ RVC 2-33 CDplus User Manual

Page 65: Calculation example

Advertising
background image

RVC 2-33 CD

&

Appendix

Version 01/2009, Rev. 1.21 of 25/11/2013 - sb

65

Translation of the original operating manual

11 Appendix

11.1 Mathematical Relations

The automatic processes under “Program administration” (see
chapter 6.3.5.2) are based on the following considerations:

Fig. 11.1:Graphical representation of the course of the control setpoint

Calculation of the control set value and of the gradient for the
temperature:

C/min]

[

sec

sec

sec

sec

tion

current

of

time

tion

tion

previous

of

value

set

tion

current

of

value

set

gradient

C]

[

sec

gradient

time

section

elapsed

tion

previous

of

value

set

value

set

control

Calculation example:

Section

Set values

Section time [h:min] Temperature [°C]

previous

...

30

current

1:00

60

C/min

0.5

min

60

30

min

60

30

60

C

C

C

gradient

After an elapsed section time of 30 minutes, for example, the control
set value for the temperature is:

C

C/

.

C

value

set

Control

t

45

min

5

0

min

30

30

min)

30

(

Calculation of the control set value for the vacuum:

]

[

10

sec

.)

sec

.

(

10

.

sec

(

10

.)

sec

.

(

10

mbar

value

set

control

time

tion

elapsed

section

current

of

time

section

t

prev

value

set

LOG

t

current

value

set

LOG

t

prev

value

set

LOG

Advertising