RVC 2-33 CD

&

Using the RVC

Version 01/2009, Rev. 1.21 of 25/11/2013 - sb

45

Translation of the original operating manual

Fig. 6.28: Entering the parameters for a pressure increase test

Enter the set values and confirm.


To perform the pressure increase test:

Select "Special functions" from the main menu.

Open the menu "Pressure increase test" and select "Start
pressure increase test". The pressure increase test will be
performed.

The test time progress and the measured pressure increase are
displayed in a graphical form.

Fig. 6.29: Graphical presentation of the pressure increase test


When the test is finished, an information window pops up and shows
the result of the pressure increase test.

6.3.5.4

Process and Equipment Information System

The "Process & Equipment Information System" informs the user
about any error, process, or system messages.


A detailed list of all messages can be found in chapter 1
"Malfunctions and Error Correction".

The process and equipment information window is displayed whenever
the CDplus control system generates a new information message. If
the user is working in a menu, the process & equipment information
window will not be displayed until the user exits the main menu.


1 Message
2 Status of the

Information

3 Number of

messages pending

4 Function key for

quitting the menu

5 Function key for

acknowledging the
message

Fig. 6.30: Process & equipment info window