Martin Christ RVC 2-33 CDplus User Manual

Page 38

Advertising
background image

RVC 2-33 CD

&

Using the RVC

38

Version 01/2009, Rev. 1.21 of 25/11/2013 - sb

Translation of the original operating manual

During the warm-up phase, the following functions can be selected
with the right-hand function key "mode":

Open lid
Continue evaporation
Stop warm-up

The warm-up is stopped by selecting one of the functions.

Fig. 6.11: Stopping the warm-up phase

6.3.4.3

Manual or Program-controlled Evaporation

If the unit is in the standby mode, the evaporation phase can be
started directly.

Press the left-hand function key to open the menu "Start with
phase…".

Select the function "Evaporation manual" or "Evaporation

program" with the up/down keys.

Press the right-hand function key to start the evaporation.

The function "Evaporation program" can only be selected if a program
was created before (see 6.3.5.2 "Program administration").

Fig. 6.12: Starting an evaporation process

6.3.4.4

Stopping the Evaporation Process

If the timer is active, the evaporation process will be stopped
automatically when the preset time is over. To abort the evaporation
process:

Press the left-hand function key. The menu "Select mode"
appears.

Select "Stop evaporation" and press the right-hand function key
to confirm.

Fig. 6.13: Stopping an evaporation process

Advertising