5 Set-up and connection

28

Version 08/2009, Rev. 2.1 of 02/12/2014 • sb

Translation of the original operating manual

3. Connection for aeration and air injection with inert gas

1 Aeration hose connection
2 Air injection hose nozzle
3 Fastener for non-required connection element; here: air filter

CAUTION

If gas (e.g. inert gas) is supplied through the hose nozzle for aeration or for
“air injection”, the threads of the hose nozzles must be sealed with thread
sealant, e.g. Loctite

®

542 (Loctite

®

part number: 54223, bottle with 10ml).

Pos: 95 / 010 Unive rsalm odul e/Le erzeil e @ 0\mo d_1 202 116 244 500 _0.d ocx @ 114 @ @ 1

Pos: 96 / 010 Unive rsalm odul e/Le erzeil e @ 0\mo d_1 202 116 244 500 _0.d ocx @ 114 @ @ 1

Pos: 97 / 200 Christ /37 0 RVC-BA (STANDARDM ODULE)/ 050 Au fstellu ng u nd Ans chluss/ 050 -00 40 V akuu mve rbind ung en (au ßer 2- 18 HCL ) @ 25\ mod _14 0498 508 723 3_6 8.do cx @ 1 870 51 @ 2 @ 1

5.4 Vacuum connections

The vacuum connection is realised by way of a vacuum hose with standard
flange connections, clamping rings, or chains, and centring rings.

NOTE

The small flange connections must be installed correctly in order to
prevent leaks (see chapter 7.1.5 - "Small flange connections ").


1

Centring ring

2

Clamping ring

3

Flange connection

Fig. 1: Connecting pieces for vacuum connection