3 overview of evaporation times – Martin Christ RVC 2-33 CDplus User Manual

Page 87

Advertising
background image

Rotational Vacuum Concentrator RVC 2-33 IR

11 Appendix

Version 08/2009, Rev. 2.1 of 02/12/2014 • sb

87

Translation of the original operating manual

Pos: 215 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /11 0 Anha ng/ 110 -003 0 Ü bersic ht Abd am pfzeit en @ 26\ mod _14 0498 534 455 6_6 8.do cx @ 1 877 79 @ 2 @ 1

11.3 Overview of evaporation times

Vessel

Number

of

samples

Sample
volume

(ml)

Temperature

(°C)

Pump

Cooling

trap

Total time

(min)

Water

10 ml test tube

24

9

45

MZ 2C

yes

635

10 ml test tube

24

9

60

MZ 2C

yes

510

1.5 ml Eppendorf caps

36

1

30

MZ 2C

yes

330

1.5 ml Eppendorf caps

36

1

45

MZ 2C

yes

270

1.5 ml Eppendorf caps

36

1

60

MZ 2C

yes

215

1.5 ml Eppendorf caps

36

1

30

KnF

no

330

1.5 ml Eppendorf caps

36

1

45

KnF

no

290

1.5 ml Eppendorf caps

36

1

60

KnF

no

230

Toluol

10 ml test tube

24

9

30

MZ 2C

yes

120

C

6

H

5

CH

3

10 ml test tube

24

9

45

MZ 2C

yes

91

10 ml test tube

24

9

60

MZ 2C

yes

65

1.5 ml Eppendorf caps

36

1

30

MZ 2C

yes

40

1.5 ml Eppendorf caps

36

1

45

MZ 2C

yes

40

1.5 ml Eppendorf caps

36

1

60

MZ 2C

yes

30

Acetonitrile

10 ml test tube

24

9

30

MZ 2C

yes

120

C

2

H

3

N

10 ml test tube

24

9

45

MZ 2C

yes

100

10 ml test tube

24

9

60

MZ 2C

yes

80

1.5 ml Eppendorf caps

36

1

30

MZ 2C

yes

45

1.5 ml Eppendorf caps

36

1

45

MZ 2C

yes

40

1.5 ml Eppendorf caps

36

1

60

MZ 2C

yes

30

1,4 Dioxane

10 ml test tube

24

9

30

MZ 2C

yes

165

C

4

H

8

O

2

10 ml test tube

24

9

45

MZ 2C

yes

110

10 ml test tube

24

9

60

MZ 2C

yes

80

1.5 ml Eppendorf caps

36

1

30

MZ 2C

yes

55

1.5 ml Eppendorf caps

36

1

45

MZ 2C

yes

45

1.5 ml Eppendorf caps

36

1

60

MZ 2C

yes

35

Advertising