Rotational Vacuum Concentrator RVC 2-33 IR

8 Maintenance and service

60

Version 08/2009, Rev. 2.1 of 02/12/2014 • sb

Translation of the original operating manual

• Remove the cable cover at the bottom of the unit.


6

Cable cover

Fig. 53: Removing the cable cover at the bottom of the unit


The cables of the electronic temperature measurement system of the shaft
can be seen now.
• Unscrew the two terminal screws of the electronic temperature

measurement system of the shaft.


7

Terminal screws

Fig. 54: Removing the terminal screws of the electronic temperature measurement system


• Turn the rotational vacuum concentrator back to its normal position

(with a sufficient number of people helping you), if it was turned upside
down.