Martin Christ RVC 2-33 CDplus User Manual

Page 50

Advertising
background image

Rotational Vacuum Concentrator RVC 2-33 IR

6 Operation

50

Version 08/2009, Rev. 2.1 of 02/12/2014 • sb

Translation of the original operating manual

Temperature Warm-up
During the warm-up phase, the rotor temperature is set to the set
temperature.

Click at button push
If this function is activated, a brief sound signal can be heard whenever a
key is pressed.

High temperature resolution
Temperatures are displayed in the values window with a resolution of

1

/

10

°C, normally with a resolution of 1°C.


Rotor speed
The rotor speed can be set in a range between 100 rpm and 1,750 rpm.

NOTE

We guarantee that a speed of 1,350 rpm can be reached with all of the
available rotors. Higher speeds are possible in individual cases.



Continuous operation coldtrap (only if a cold trap is installed) and
Continuous operation VP
If these options are set to "yes", the vacuum pump and the cold trap will not
be switched off after the evaporation. Instead, they will remain active. This
is useful if several evaporation processes are started in succession.

Service menu

Fig. 44: Service menu


Rotor information
This function is not supported by this type of unit.