7 Malfunctions and error correction

Version 08/2009, Rev. 2.1 of 02/12/2014 • sb

53

Translation of the original operating manual

Error

Possible reason

Correction

Permissible imbalance exceeded

• The rotor will be stopped

immediately, the process will

also be stopped.
– improper loading of the

rotor

– glass breakage
– uneven evaporation

• Balance the load and restart the

rotational vacuum concentrator.
If the error occurs again, contact
the after-sales service (see

chapter 7.2 - "Service contact")

VSK3000 calibration error

• Incorrect calibration of the

vacuum sensor VSK3000

• The sensor must be recalibrated

(see separate operating manual)

Pos: 158 /01 0 Univ ersal mod ule/L eerz eile @ 0\m od_ 120 211 624 450 0_0. docx @ 11 4 @ @ 1

Pos: 159 /01 0 Univ ersal mod ule/L eerz eile @ 0\m od_ 120 211 624 450 0_0. docx @ 11 4 @ @ 1

Pos: 160 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /07 0 Stör ung en u nd Fehl ersuc he/ 070 -00 10- 002 0 Pro zess meldu nge n ( auß er 2-1 8) @ 25\ mod _14 049 851 884 95_6 8.d ocx @ 187 457 @ 3 @ 1

7.1.2 Process messages

Error

Possible reason

Correction

Mains power failure

• A mains power failure will be

detected if the power supply is
interrupted under normal

conditions (run mode).

• The process will continue once

the mains power is available
again. The process times will be
reset to zero. The set control

values will be preserved.

Factory settings loaded

• All parameters are reset to the

delivery status of the system.

• The factory settings will be

loaded if the corresponding
menu command is executed in
the service menu or if an error

occurred in the parameter menu.

No rotor detected (only Triple TMM)

• Check the rotor for correct

installation.

• If the error occurs again, contact

the after-sales service (see
chapter 7.2 - "Service contact").

Nominal rotor speed not reached

• After two minutes, the rotor has

still not reached 90% of its

nominal speed.
– set speed is too high
– drive moves sluggishly

• Check the set speed.
• When the rotor is at a standstill

and the system is switched off,
check whether the drive moves
sluggishly, by hand. Contact the
after-sales service (see chapter

7.2 - "Service contact").

Safety pressure not reached

• The safety pressure inside the

rotor chamber must be reached
within 5 minutes after the start

of an evaporation process.

• Check whether the value that

was entered for the safety

pressure is not too low.

• Check the components for any

possible leaks.

Ice condenser temperature not

reached

• Only if the "Cooling trap control"

option is enabled:
The ice condenser temperature
must be

-25°C.

• If the temperature is not

reached, the heater will be

deactivated.