Rotational Vacuum Concentrator RVC 2-33 IR

8 Maintenance and service

Version 08/2009, Rev. 2.1 of 02/12/2014 • sb

61

Translation of the original operating manual

• Remove the cap.

Fig. 55: Removing the cap


• Screw out the electronic temperature measurement system of the

shaft. Pay attention to the corresponding cables.

Fig. 56: Removing the electronic temperature measurement system


• Loosen the rotor shaft incl. magnetic coupling using the hexagon socket

wrench (size 5).

Fig. 57: Loosening the rotor shaft incl. magnetic coupling