Martin Christ RVC 2-33 CDplus User Manual

Page 63

Advertising
background image

Rotational Vacuum Concentrator RVC 2-33 IR

8 Maintenance and service

Version 08/2009, Rev. 2.1 of 02/12/2014 • sb

63

Translation of the original operating manual

• Clean the cap and grease the O-ring of the cap slightly with vacuum

grease.

• Install the components or assemblies in a reversed order.

CAUTION

When installing the rotor shaft, ensure that the O-ring used to seal the
rotor chamber is fitted correctly.


• Grease the O-ring with vacuum grease to ensure that it does not get out

of place.

8

O-Ring of the rotor

shaft

Fig. 60: O-ring at the bottom of the rotor shaft


• Plug in the mains plug and switch the rotational vacuum concentrator

on.

• Perform a test run.

Pos: 178 /01 0 Univ ersal mod ule/L eerz eile @ 0\m od_ 120 211 624 450 0_0. docx @ 11 4 @ @ 1

Pos: 179 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /08 0 Wart ung un d Inst and haltu ng/ 080 -001 0- 0030 Z ube hör @ 2 5\m od_ 1404 985 237 565 _68. docx @ 18 755 5 @ 3 @ 1