Rotational Vacuum Concentrator RVC 2-33 IR

5 Set-up and connection

Version 08/2009, Rev. 2.1 of 02/12/2014 • sb

31

Translation of the original operating manual

Pos: 105 /20 0 Ch rist/3 71 RVC-BA (PROJEK TE)/RVC 2- 33 IR /05 0 Aufst ellung und Anschlus s/05 0-0 060 -00 20 K ond ensati on d er D ämp fe ü ber Kühlfall e RVC 2- 33 IR @ 31 \mo d_1 407 484 067 433_ 68. docx @ 20 907 0 @ 3 @ 1

5.6.2

Condensation of the vapours upstream of the vacuum pump in a cooling
trap

The vapours are condensed upstream of the vacuum pump in a cooling
trap, e.g.”CT 02-50” or “CT 04-50”.
The application is suitable for water-base, low-boiling samples containing
solvents. The RVC, the cooling trap and the vacuum pump must be
connected. The connector of the stop valve must be plugged into the
socket on the back of the unit.

1

RVC

2

Vacuum sensor

3

Stop valve

4

Cover with

connecting hoses

5

Cooling trap

6

Pressure control

valve

7

Vacuum pump

Fig. 10: Combination of the RVC with vacuum pump and cooling trap

Pos: 106 /01 0 Univ ersal mod ule/L eerz eile @ 0\m od_ 120 211 624 450 0_0. docx @ 11 4 @ @ 1

Pos: 107 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /05 0 Aufstell ung und Anschlus s/05 0-0 060 -00 20- 001 0 O ption : F ern bedie nun g d er Kühl falle mit d em RVC @ 25\ mod _14 049 850 892 21_6 8.d ocx @ 187 079 @ 4 @ 1

5.6.2.1

Option: Remote control of the cooling trap via RVC

If CHRIST freeze dryers are used as a cooling trap, they can be remote-
controlled by the rotational vacuum concentrator. In this case, the cooling
trap will be automatically activated by the rotational vacuum concentrator
during the warm-up and evaporation phase. In the standby mode, the
cooling trap will be either switched off or continue operation, depending on
the choosen option (chapter 6.5.3.5 - "Options" Settings / Continuous
operation coldtrap"). The ice condenser temperature is indicated in the
values window of the rotational vacuum concentrator.

The cooling trap must be connected to the rotational vacuum concentrator
with the aid of a zero cross cable (part no. 222 000). In addition, the option
"Cooling trap control" must be activated (see chapter 6.5.3.5 - "Options").